Hospitals and neonatal units have been instructed to stop using and destroy certain Portex endotracheal tubes after a manufacturing issue made some tubes softer than intended, according to an FDA early alert posted Nov. 25. “As of November 16, 2026, ICU Medical has reported one death associated with this issue,” the FDA said.
The affected products are small, uncuffed breathing tubes in 2.5 mm, 3.0 mm, and 3.5 mm sizes. The FDA warning focuses on newborns and young infants, whose airways are tiny and whose breathing tubes leave little room for error.
Parents will not handle these tubes directly. However, families with a baby in a neonatal intensive care unit or a newborn who may need surgery may wonder whether their child’s care is affected. The alert is directed at hospitals, which are responsible for checking their supplies.
Item Numbers and the Plasticizer Problem
The alert covers five Portex products made by ICU Medical: uncuffed Murphy eye tracheal tubes for oral or nasal use, item numbers 100/127/025, 100/127/030, and 100/127/035, and Polar Clear South oral uncuffed tracheal tubes, item numbers 100/134/030 and 100/134/035. Affected lots are listed in a separate document linked from the FDA page.
ICU Medical found that the resin used in certain lots contained varying amounts of plasticizer, a substance that makes plastic flexible. That variation can leave a tube softer than intended.
An endotracheal tube is placed through the mouth or nose into the windpipe so a ventilator or clinician can deliver air to the lungs, according to MedlinePlus. In newborns, the tube must be firm enough to guide into place and stay open, yet small enough for an airway only a few millimeters wide.
The FDA says that in newborns and young infants, the softness can cause tube kinking and delayed intubation because of increased resistance to ventilation. If the problem is not identified during intubation, it may lead to inadequate breathing support and unplanned tube changes, which “may lead and/or contribute to serious adverse events including death.”
Instructions for Hospitals and NICUs
ICU Medical sent customers a letter on Nov. 20. It asks facilities to check all inventory locations, stop using affected tubes, and destroy them according to institutional procedures. If destruction cannot be completed immediately, products should be quarantined so they cannot be used.
Facilities are also asked to post the company’s notice near affected stock until it is destroyed, forward the notice to anyone to whom they distributed the tubes, and report adverse events to ICU Medical’s complaint line at 866-216-8806. Customer service is available at 800-258-5361.
The FDA issued the notice as an early alert, which it uses to flag a potentially high-risk device problem before a full recall. The agency said it will update the page as significant new information becomes available.
Removing common sizes could strain supply, and the FDA has not said whether alternatives are in short supply. Neonatal units, delivery rooms, emergency departments, and transport teams all stock small tubes, so hospitals may need to confirm they have enough alternatives before pulling inventory.
Questions Families Can Ask
The babies most likely to need these small tubes are premature infants, newborns with breathing problems at birth, and infants undergoing surgery.
Neonatal intubation is one of the hardest procedures in medicine. Clinicians often have only seconds to place a tube in a struggling newborn, so a tube that bends or resists, even slightly, can cost precious time. That is why a material defect matters more here than it might in adult care.
Parents of a baby who is or was recently intubated can ask the care team whether the hospital used any of the affected products and whether the alert has been addressed. The question is reasonable, not accusatory.
A baby who was intubated without problems and is now breathing well does not need extra testing because of this alert. Families who have concerns about a past procedure, such as a difficult intubation or unexpected tube change, can ask for a review with the medical team.
MedicalDaily Evidence Check: This is an FDA early alert based on information reported by the manufacturer. The single death was reported by ICU Medical to the FDA, and the alert does not describe the circumstances or report a count of injuries. The FDA has not yet confirmed a recall classification.
Reporting and Next Steps
Health care workers and families can report problems with any medical device to the FDA’s MedWatch program. Reports help the agency identify patterns, especially for devices used on vulnerable patients.
The FDA keeps a running list on its recalls and early alerts page, which hospital supply teams and patient safety officers can use to check inventory. Portex breathing tubes have been flagged before: Smiths Medical, the brand’s former owner, issued an earlier correction for some Portex tubes because certain sizes were smaller than expected.
Several details remain unknown. The alert does not say how many tubes were distributed, when the affected lots were shipped, or whether the death occurred in the U.S. ICU Medical has not publicly described the case. Those gaps make it hard to judge how widespread the problem is.
MedicalDaily will report any recall classification, injury reports, or expansion of affected lots.
Key Questions Answered
What is being removed? Certain lots of five Portex uncuffed endotracheal tubes in 2.5 mm, 3.0 mm, and 3.5 mm sizes made by ICU Medical.
What is wrong with them? A varying plasticizer level can make the tubes too soft, causing kinks and delays in placing them.
Has anyone been hurt? ICU Medical reported one death associated with the issue as of Nov. 16, according to the FDA.
Who uses these tubes? Mainly newborns and young infants who need breathing support or surgery.
Should parents be worried? Hospitals are responsible for removing the tubes. Parents can ask the care team whether the alert has been addressed.
How can problems be reported? To ICU Medical at 866-216-8806 or the FDA’s MedWatch program.
Also Read
- Experimental Drug PHI-101 Doubles Survival in Mice With Drug-Resistant AML, Johns Hopkins Study Finds
- Low-Oxygen Therapy Triples Lifespan in Mice with Mitochondrial Brain Disorder
- Newborn’s Arm Swelling After NICU Discharge Tied to Leaked IV Calcium
- New Identity Verification Rules Stall Affordable Care Act Enrollment in Texas as the Dec. 15 Deadline Nears